Gastec Tubes and Sampling Pump
Short terms quick measuring detector tubes 
  A  
  B  
  C  
  D  
  E  
  F  
  G  
  H  
  I  
  L  
  M  
  N  
  O  
  P  
  S  
  T  
  V  
  W  
  X  
Gas sampling pump kit and accessories
 
 
 
 
Acetaldehyde  

NO.92L

CH3CHO
Performance  
Measuring Range
1 to 20 ppm
Number of Pump Strokes
1
Correction Factor
1
Sampling Time 3 minutes per pump stroke
Detecting Limit 0.2 ppm (n=1)
Colour Change Yellow → Brown
Reaction Principle 3CH3CHO + (NH2OH)3H3PO4 → H3PO4
H3PO4 + Base → Phosphate
Coefficient of Variation 10% (for 1 to 6 ppm), 5% (for 6 to 20 ppm)
Shelf Life 2 Years
Corrections for temperature & humidity Temperature correction is necessary
@
 
Possible coexisting substances and their interferences (NOTE)
Substance Concentration Interference Change colour by itself to
Ammonia 30 ppm No No
Acetic acid 2 times + Brown
Acids   + Brown
Aldehydes   + Brown
Ketones   + Brown
Calibration gas generation Diffusion tube method
TLV-TWA TLV-STEL Explosive range
- C 25 ppm@ 4 to 60%
 
 
   
 

2008, RespoRae Systems. All rights reserved.