Gastec Tubes and Sampling Pump
Short terms quick measuring detector tubes 
  A  
  B  
  C  
  D  
  E  
  F  
  G  
  H  
  I  
  L  
  M  
  N  
  O  
  P  
  S  
  T  
  V  
  W  
  X  
Gas sampling pump kit and accessories
 
 
 
Formaldehyde   NO.91
HCHO
Performance  
Measuring Range
2 to 20 ppm 20 to 50 ppm 50 to 100 ppm
Number of Pump Strokes
2 1 1/2
Correction Factor
1 2.5 5
Sampling Time 1 minute per pump stroke
Detecting Limit 0.5 ppm (n=2)
Colour Change White → Brown
Reaction Principle HCHO + C6H4(CH3)2 + H2S2O7 → Condensation Polymer
Coefficient of Variation 15% (for 2 to 6 ppm), 10% (for 6 to 20 ppm)
Shelf Life 3 Years
Corrections for temperature & humidity Unnecessary
Store the tubes at cool and dark place.
 
Possible coexisting substances and their interferences (NOTE)
Substance Concentration Interference Change colour by itself to
Aldehydes 1/2 time + Brown
Esters, Ethers 1000 ppm + No
Ketones 1000 ppm + No
Styrene 500 ppm + Yellow
Calibration gas generation Diffusion tube method
TLV-TWA TLV-STEL Explosive range
- C 0.3 ppm 7.0 to 73%
 
 
 
   
 