You are Here: » Short
Term Tubes » A, B, F, M, P & V » 81L |
| |
|
|
|
Acetic acid |
|
NO.81L |
|
CH3CO2H |
 |
|
Performance |
|
|
Measuring Range |
|
0.125 to 0.25 ppm |
0.25 to 10 ppm |
10 to 25 ppm |
|
|
Number of Pump Strokes |
|
|
Correction Factor |
|
| Sampling
Time |
1.5
minutes per pump stroke |
| Detecting
Limit |
0.05 ppm
(n=2) |
| Colour
Change |
Pink →
Pale Yellow |
| Reaction
Principle |
CH3CO2H
+ Base → Reaction product |
|
Coefficient of Variation |
10% (for
0.25 to 3 ppm), 5% (for 3 to
10 ppm) |
| Shelf
Life |
2 Years |
|
Corrections for temperature
& humidity |
Temperature correction is
necessary |
|
Store the tubes at in the
refrigerator to keep at
10ºC(50ºF) or below. |
|
|
|
Possible coexisting substances and
their interferences
(NOTE) |
|
Substance |
Concentration |
Interference |
Change colour by itself to |
|
Ammonia |
Higher than 2 times |
|
No |
|
Chlorine, sulphur dioxide,
Nitrogen dioxide |
|
+ |
Pale Yellow |
|
Formic Acid, Acetic
anhydride |
|
+ |
Pale Yellow |
|
|
|
|
|
|
Other substance measurable with this
detector tube |
|
Substance |
Correction Factor |
No. of pump strokes |
Measuring Range |
|
Acetic anhydride |
Factor: 0.6 |
1 |
0.15 to 6 ppm |
|
Acrylic acid |
Factor: 1.8 |
1 |
0.45 to 18 ppm |
|
Butyric acid |
Factor:1.3 |
1 |
0.325 to 13 ppm |
|
Formic acid |
Factor: 2.0 |
1 |
0.5 to 20 ppm |
|
Isovaleric acid |
Factor: 1.5 |
1 |
0.38 to 15 ppm |
|
Methacrylic acid |
Factor 1.4 |
1 |
0.35 to 14 ppm |
|
Propionic acid |
Factor: 1.0 |
1 |
0.25 to 10 ppm |
|
Valeric acid |
Factor: 1.5 |
1 |
0.38 to 15 ppm |
|
Calibration gas generation |
Diffusion tube method |
|
TLV-TWA |
TLV-STEL |
Explosive range |
|
- |