You are Here: Short
Term Tubes A, B, M & P 191L |
| |
|
|
|
Acrylonitrile |
|
NO.191L |
|
CH2FCHCN |
 |
|
Performance |
|
|
Measuring Range |
|
0.1 to 0.2 ppm |
0.2 to 6.0 ppm |
6.0 to 18.0 ppm |
|
|
Number of Pump Strokes |
|
|
Correction Factor |
|
| Sampling
Time |
3 minutes
per pump stroke |
| Detecting
Limit |
0.05 ppm
(n=4) |
| Colour
Change |
Yellow →
Pink |
| Reaction
Principle |
Pretreatment tube: CH2:CHCN
+ Cr6+ + H2SO4
→ HCN
Detector tube : 2HCN +
HgCl2
→ 2HCl + Hg(CN)2
HCl + Base →
Chloride |
|
Coefficient of Variation |
10% (for
0.25 to 1 ppm), 5% (for 1 to
6 ppm) |
| Shelf
Life |
3 Years |
|
Corrections for temperature
& humidity |
Unnecessary |
|
Store the tubes at cool and
dark place. |
|
|
|
Possible coexisting substances and
their interferences
(NOTE) |
|
Substance |
Concentration |
Interference |
Change colour by itself to |
|
Hydrogen chloride |
|
No |
No |
|
Hydrogen cyanide |
|
No |
No |
Nitriles ( C3) |
|
+ |
Pink |
|
Acetone cyanohidrin |
|
+ |
Pink |
|
Alcohols, Esters, Ketones |
|
No |
No |
|
Aromatic hydrocarbons |
|
No |
No |
|
Other substance measurable with this
detector tube |
|
Substance |
Correction Factor |
No. of pump strokes |
Measuring Range |
|
n-Butyronitrile |
30 |
1 |
6 to 180 ppm |
|
2-Methyl-3-butenenitrile |
2.0 |
2 |
0.4 to 12 ppm |
|
2-Pentenenitrile |
1.2 |
2 |
0.24 to 7.2 ppm |
|
3-Pentenenitrile |
2.0 |
2 |
0.4 to 12 ppm |
|
Calibration gas generation |
Diffusion tube method |
|
TLV-TWA |
TLV-STEL |
Explosive range |
|
2 ppm |
- |
3 to 17% |
|
|
|
|
|
|
|