You are Here: » Short
Term Tubes » D, E & P » 163 |
| |
|
|
|
Ethylene oxide |
|
NO.163 |
|
C2H4O |
 |
|
Performance |
|
|
Measuring Range |
(0.05 ) to 3% |
|
Number of Pump
Strokes |
1 |
|
Correction Factor |
1 |
|
Sampling Time |
1
minute per pump
stroke |
|
Detecting Limit |
0.02% (n = 1) |
|
Colour Change |
Orange → Green |
|
Reaction Principle |
C2H4O+
Cr6+ + H2SO4
→ Cr3+ |
|
Coefficient of
Variation |
10% (for 0.05 to 1.0
%), 5% (for 1.0 to
3.0 %) |
|
Shelf Life |
3
Years |
|
Corrections for
temperature &
humidity |
Unnecessary |
|
Store the tubes at
cool and dark place. |
|
|
Possible coexisting substances and
their interferences
(NOTE) |
|
Substance |
Concentration |
Interference |
Change colour by itself to |
|
Aromatic HCs |
|
+ |
Green |
|
Alcohols, Esters,
Ethers |
|
+ |
Green |
|
|
Other substance measurable with this
detector tube |
|
Substance |
Correction Factor |
No. of pump strokes |
Measuring Range |
|
1,4-Dioxane |
Factor: 2.0 |
1 |
0.1 to 6% |
|
Propylene oxide |
Factor: 1.3 |
1 |
0.065 to 3.9% |
|
|
Calibration gas generation |
Static gas dilution method |
|
TLV-TWA |
TLV-STEL |
Explosive range |
|
1 ppm |
- |
3.6 to 100% |
|
|
|
|
|
|
|