Gastec Tubes and Sampling Pump
Short terms quick measuring detector tubes 
  A  
  B  
  C  
  D  
  E  
  F  
  G  
  H  
  I  
  L  
  M  
  N  
  O  
  P  
  S  
  T  
  V  
  W  
  X  
Gas sampling pump kit and accessories
 
 
 
Hydrogen cyanide  

NO.12LL

HCN
Performance
Measuring Range 0.2 to 7 ppm
Number of Pump Strokes 2
Correction Factor 1
Sampling Time 2 minutes per pump stroke
Detecting Limit 0.05 ppm (n=2)
Colour Change Yellow → Pink
Reaction Principle 2HCN + HgCl2 → Hg(CN)2 + 2HCl
HCl + Base → Chloride product
Coefficient of Variation 10% (for 0.2 to 2 ppm), 5% (for 2 to 7 ppm)
Shelf Life 2 Years
Corrections for temperature & humidity Humidity correction is necessary
Store the tubes at cool and dark place.
 
Possible coexisting substances and their interferences (NOTE)
Substance Concentration Interference Change colour by itself to
Hydrogen sulphide 1 ppm + Red
sulphur dioxide 1 ppm + Red
Calibration gas generation Permeation tube method
TLV-TWA TLV-STEL Explosive range
- C 4.7 ppm 5.6 to 40%
 
 
 
   
 

© 2008, RespoRae Systems. All rights reserved.