Gastec Tubes and Sampling Pump
Short terms quick measuring detector tubes 
  A  
  B  
  C  
  D  
  E  
  F  
  G  
  H  
  I  
  L  
  M  
  N  
  O  
  P  
  S  
  T  
  V  
  W  
  X  
Gas sampling pump kit and accessories
 
 
 
Hydrocarbons (Higher Class)  

NO.105

C5`C10
Performance
Measuring Range
100 to 200 ppm 200 to 3000 ppm
Number of Pump Strokes
2 1
Correction Factor
1/2 1
Sampling Time 1.5 minutes per pump stroke
Detecting Limit 20 ppm (n=2)
Colour Change White → Blackish brown
Reaction Principle Hydorocarbon (C6 + C10) + I2H5 + H2S2O7 → I2
Coefficient of Variation 10% (for 200 to 1000 ppm), 5% (for 1000 to 3000 ppm)
Shelf Life 3 Years
Corrections for temperature & humidity Unnecessary
Store the tubes at cool and dark place.
 
Possible coexisting substances and their interferences (NOTE)
Substance Concentration Interference Change colour by itself to
Acetylene, Ethylene 0.1% + Blackish brown
Carbon monoxide 0.1% + Blackish brown
Other organic solvents   + Blackish brown
 
Other substance measurable with this detector tube
Substance Correction Factor No. of pump strokes Measuring Range
Decane 2.0 1 or 2 200 to 6000 ppm
Heptane 0.9 1 or 2 90 to 2700 ppm
Hexane 0.8 1 or 2 80 to 2400 ppm
Nonane 1.3 1 or 2 130 to 3900 ppm
Octane 1.0 1 or 2 100 to 3000 ppm
Calibration gas generation Vaper pressure method
TLV-TWA TLV-STEL Explosive range
- - -
 
 
   
 